TGuide Cells/Tissue/Plant RNA Kit

For extracting total RNA from samples of cells, tissues, plants, etc.

TGuide Cells/Tissue/Plant RNA Kit is specially designed to extract high purity RNA from animal cells, animal tissues and plant tissues using TGuide Series Automated Nucleic Acid Extractor, with no contamination of protein and other impurities. The kit contains reagents and consumables required for automatic DNA extraction by magnetic bead method. The reagents are pre-packed in sealed reagent cartridges. The unique embedded magnetic beads and fully automatic extraction process ensure the quick and convenient RNA purification.

Cat. No Packing Size
OSR-M610 48 preps

Product Detail

Product Tags

Applications

The purified RNA can be directly used in quantitative RT-PCR, RTPCR, cDNA synthesis and other experiments.

Features

■ Simple and fast extraction: TGuide accessory products are based on the principle of nucleic acid purification by magnetic beads and RNA extraction process can be completed within 72 min.
■ No cotamination: Independent sealed reagent cartridge and consumables with DNase/RNase removal treatment to effectively reduce RNase contamination and cross-contamination.

All the products can be customized for ODM/OEM. For details, please click Customized Service(ODM/OEM)

FAQ

Q: Column blockage

A-1 Cell lysis or homogenization not sufficient

---- Reduce sample usage, increase the amount of lysis buffer, increase homogenization and lysis time.

A-2 Sample amount is too large

---- Reduce the amount of sample used or increase the amount of lysis buffer.

Q: Low RNA yield

A-1  Insufficient cell lysis or homogenization

---- Reduce sample usage, increase the amount of lysis buffer, increase homogenization and lysis time.

A-2  Sample amount is too large

----Please refer to the maximum processing capacity.

A-3 RNA is not eluted completely from the column

---- After adding RNase-Free water, leave it for a few minutes before centrifuging.

A-4 Ethanol in the eluent

---- After rinsing, centrifuge again and remove the washing buffer as much as possible.

A-5 Cell culture medium is not completely removed

---- When collecting cells, please make sure to remove the culture medium as much as possible.

A-6 The cells stored in RNAstore are not effectively centrifuged

----RNAstore density is greater than the average cell culture medium; so the centrifugal force should be increased. It is suggested to centrifuge at 3000x g.

A-7 Low RNA content and abundance in the sample

---- Use a positive sample to determine if the low-yield is caused by the sample.

Q: RNA degradation

A-1  The material is not fresh

---- Fresh tissues should be stored in liquid nitrogen immediately or immediately put into the RNAstore reagent to ensure the extraction effect.

A-2  Sample amount is too large

---- Reduce sample amount.

A-3  RNase contamination

----Although the buffer provided in the kit does not contain RNase, it is easy to contaminate RNase during extraction process and should be handled with care.

A-4  Electrophoresis pollution

---- Replace the electrophoresis buffer and make sure the consumables and Loading Buffer are free of RNase contamination.

A-5  Too much loading for electrophoresis

---- Reduce the amount of sample loading, the loading of each well should not exceed 2 μg.

Q: DNA contamination

A-1  Sample amount is too large

---- Reduce sample amount.

A-2  Some samples have high DNA content and can be treated with DNase.

----Perform RNase-Free DNase treatment to the obtained RNA solution, and the RNA can be directly used for subsequent experiments after treatment, or can be further purified by RNA purification kits.

Q: How to remove RNase from experimental consumables and glasswares?

For glasswares, baked at 150°C for 4 h. For plastic containers, immersed in 0.5 M NaOH for 10 min, then thoroughly rinsed with RNase-free water and then sterilize to completely remove RNase. The reagents or solutions used in the experiment, especially water, must be free of RNase. Use RNase-free water for all reagent preparations (add water to a clean glass bottle, add DEPC to a final concentration of 0.1% (V/V), shake overnight and autoclave).